리퍼비시 계측기

리퍼비시 계측기

ZV-Z235

  • ModelZV-Z235
  • MakerR&S
  • DescriptionCalibration Kits, 0 Hz to 24 GHz , 50 Ω , 3.5 mm female and male
  • Rental문의요망
  • Stock문의요망

The specified effective system data are established after performing a suitable system error calibration, e.g. TOSM, at a R&S®
ZVA,
R&S®
ZVB, or R&S®
ZVT vector network analyzer, using the characteristic data of the calibration kit, which are stored on a provided
USB stick. This data is valid between +18 °C and +28 °C, at a measurement bandwidth of 10 Hz, and a nominal power of −10 dBm at
the calibration ports. The calibration kit is fully functional down to 0 Hz, with effective system data as specified below, although the
data is only verified for frequencies as stated.

Directivity 10 MHz to 700 MHz > 36 dB, typ. 46 dB
                      700 MHz to 24 GHz > 40 dB, typ. 46 dB
Source match 10 MHz to 700 MHz > 30 dB, typ. 43 dB
                           700 MHz to 24 GHz > 36 dB, typ. 43 dB
Reflection tracking 10 MHz to 700 MHz < 0.2 dB, typ. 0.04 dB
                                   700 MHz to 24 GHz < 0.1 dB, typ. 0.02 dB
Load match 10 MHz to 700 MHz > 36 dB, typ. 46 dB
                       700 MHz to 24 GHz > 40 dB, typ. 46 dB
Transmission tracking 10 MHz to 700 MHz < 0.2 dB, typ. 0.04 dB
                                           700 MHz to 24 GHz < 0.1 dB, typ. 0.02 dB